We are a noted firm involved in manufacturing, supplying and exporting Evita 4 Dura Ventilator. Irays Medtech Suppliers, Dealer & Importer of high quality Refurbished Drager Evita 4 Medical Ventilator, Refurbished EVITA Ventilator, Drager Evita 4/XL Respiratory Ventilators, Respiratory Ventilators. Our Refurbished Drager EVITA Ventilator was customized to meet the accepting requirements of the ICU environment and Drager EVITA 4 is a Ventilator for every patient rush by improving the communications between patient, ventilator and clinician. The functional touch monitor frequently provides the clinician information on ventilator settings, patient measurements and advanced trending capabilities which enhances the operation of the device. Unique features and modalities such as AutoFlow™ and APRV are standard on all Evita 4 ventilators.

Technical Specifications

Ventilation settings:

Ventilation mode:

  • IPPV, IPPVAssist (CMV, CMVAssist)
  • SIMV, SIMVASB (SIMV, SIMV/Psupp)
  • MMV, MMVASB (MMV, MMV/Psupp)
  • BIPAP1), BIPAP1)ASB, BIPAP1)Assist (PCV+, PCV+/Psupp, PCV+Assist)
  • APRV
  • CPAP, CPAPASB (CPAP, CPAP/Psupp)
  • ILV
  • PPS (optional)

Enhancements:

  • AutoFlowTM – Automatic adaptation of inspiratory flow in volume controlled modes
  • ATCTM – Automatic Tube Compensation (optional)
  • NIV – Mask Ventilation (optional)
  • Ventilation frequency (f): 0 to 100 /min, 0 to 150 /min (Neonatal)
  • Inspiration time (Tinsp: 0.1 to 10 s
  • Tidal volume (VT) (BTPS): 0.1 to 2.0 L (Adult), 0.02 to 0.3 L (Pediatric), 0.003 to 0.1 L (Neonatal)
  • Inspiratory flow: 6 to 120 L/min (Adult), 6 to 30 L/min (Pediatric and Neonatal)
  • Inspiratory pressure: 0 to 80 mbar (cmH2O)
  • PEEP / intermittent PEEP: 0 to 35 mbar (cmH2O)
  • Pressureassist (PASB) (Psupp): 0 to 80 mbar (cmH2O)
  • Rise time for inspiratory pressure: 0 to 2 s
  • O2 concentration: 21 to 100 Vol.%
  • Trigger sensitivity: 0.3 to 15 L/min

Measured values displayed:

  • Airway pressure: Peak pressure, plateau pressure, mean pressure, PEEP, min. pressure (0 to 99 mbar/cmH2O)
  • Minute volume (MV), (BTPS): MV, MVspont, MVleak (0 to 99 L/min)
  • Tidal volume (VT), (BTPS): Inspired VT, expired VT, VTPS (0 to 3999 mL)
  • Breathing frequency (f): ftotal, fspon, fmand. (0 to150/bpm)
  • O2 concentration (FiO2): Inspired O2 concentration (15 to 100 Vol.%)
  • Breathing gas temperature: 18° to 51°C
  • Waveforms: Airway pressure-time, flow-time, volume-time
  • Trends (8 anyone configurable): FiO2, MV, VT, f, PEEPi, R, C, etCO2
  • Capnography (etCO2) (optional): 0 to 100 mmHg
  • CO2 production (VCO2): 0 to 999 mL/min, STPD*
  • Serial dead space Vds: 0 to 999 mL, BTPS*
  • Dead space ventilation (Vds/VT): 0 to 99 %
  • O2 saturation: SpO2, pulse

Alarms / Monitoring:

  • Airway pressure: High / Low
  • Expired minute volume: High / Low
  • Tidal volume: High
  • Apnea alarm time: 5 to 60 s
  • Spontaneous breath frequency: High
  • Inspired O2 concentration: High / Low
  • Breathing gas temperature: High
  • SpO2 pulse (optional): High / Low
  • etCO2 (optional): High / Low

Performance data:

  • Max. flow for pressure support and spontaneous breathing: 180 L/min (adult), 60 L/min (pediatric)
  • Valve response time: T0...90 ≤ 5 ms
  • Control principle: Time cycled, volume constant, pressure-controlled
  • Safety relief valve: 100 mbar (cmH2O)